Ventilation control

Product Center
pal Lab focuses on laboratory EPC engineering and is your ideal laboratory partner.

Laboratory negative pressure control system

Room negative pressure control solution

improve air circulation Air entering the laboratory from non laboratory environmental areas should be directly discharged from the outside of the laboratory building

-Occupational Safety and Health Administration (OSHA)

Under any conditions, the air entering the laboratory area should come from areas such as office areas, corridors, etc

-Prudent Practices

In the process of indoor pressure control in the laboratory, maintaining a specific pressure difference between indoor and outdoor is not our ultimate goal,

What we really need to achieve is to maintain a constant flow of air at a certain speed when circulating indoors and outdoors“

-American National Standards Institute (ANSI)/American Industrial Hygiene Association (AIHA) Z9.5

There are three ways to control negative pressure:

1. Pressure difference control method:

The laboratory should always maintain negative pressure

Ensure that there is always a certain pressure difference between the laboratory and the corridor

2. Residual air control method:

As shown in the example diagram, ensure that the residual air volume is 10% * total exhaust air volume


3. Compensation control method:

Residual air volume control+pressure difference control

Control through residual air flow, adjust the residual air flow value when the pressure difference deviates